An internal standard is a known chemical added to a sample to calibrate laboratory instruments, allowing for the calculation of unknown quantities. Gas and liquid chromatography separate samples into different molecules, and adding an internal standard provides a known result for any sample placed into the test instrument. The relative amount of the unknown chemical can be determined by knowing the amount of internal standard, allowing for a concentration calculation. The known standard will always remain the same from sample to sample, regardless of other conditions that may change.
An internal standard is a known amount of chemical added to an analytical sample, which allows technicians to calibrate laboratory instruments. This also allows calculation of unknown chemical quantities, as the instrument will show the quantity of the known standard as a reference. Tests that can benefit from an internal standard include gas and liquid chromatography.
Gas chromatography passes a vaporized sample through a column filled with a very pure, solid form of medium. The sample is moved through the column with an inert or unreactive gas such as nitrogen or argon. Different molecules absorb onto the column and then are released according to their molecular structure. The gas then carries the separated sample to a flame to create ions or molecules with small electrical charges. A detector determines the amount of each ion and plots the results on a graph.
Liquid chromatography also separates samples into different molecules, but uses a liquid solvent to move the sample by gravity or pressure through a stationary material. A chromatography bed material can be a resin, diatomaceous earth, or even some other liquid. The sample is loaded to the top and solvent is added to push the sample down through the column material.
Parts of the sample will pass through the column at different rates based on molecular or ionic characteristics. Small samples of liquid, called aliquots, are taken at regular intervals from the bottom of the column. These samples will contain the different molecules that have separated in the column and will be analyzed to determine their composition.
In both types of chromatography, many variables can affect the results. The carrier gas flow rate may vary slightly or the columns may be at different temperatures for different samples. Adding a calibrated internal standard provides a known result for any sample placed into the test instrument. If some variable in the instrument, such as gas flow rate, changes by a small percentage, the standard will change by the same amount.
Once the instrument has been calibrated by running a sample with the internal standard, an unknown chemical sample can be run through the same instrument, with the standard included. The results will show the chemical composition of the unknown, plus the internal standard, plotted on the graph created by the instrument. By knowing the amount of internal standard, the relative amount of the unknown chemical can be determined, which will allow for a concentration calculation.
It is important to note that an internal standard is not used to chemically react with any sample. Rather, it is used to compare molecules of any unknown material to the known amount of the standard. This technique is known as setting a variable, because the known standard will always remain the same from sample to sample, regardless of other conditions that may change.